6921 Epicardial Patch
Defibrillation
6921 Epicardial Patch
ALL
Information

Device Survival Probability


6721_6921_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 111 mo
% Survival 96.5 95.0 92.7 91.9 90.0 85.1 83.7 83.7 83.7 83.7
#   330 301 256 209 176 132 95 66 56 50

System Longevity Study Results

Number of Leads Enrolled in Study 407
Cumulative Months of Follow-Up 23303
Number of Leads Active in Study 4
Information 47

Qualifying Complications

Conductor Fracture 21
Failure to Capture 8
Impedance Out of Range 4
Insulation (not further defined) 2
Oversensing 12
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

Information

US Acute Lead Observations

Information

Distribution Data

US Market Release1993-02-11
CE Approval Date
Registered USA Implants5978
Estimated Active USA Implants409
US Market Release1993-02-11
CE Approval Date
Estimated WW Distribution6187
Information

USA Returned Product Analysis

Conductor Fracture 13
Insulation Breach 1
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 14
Insulation Breach 1
Crimp/Weld/Bond 0
Other 0
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Suture
Lead Function Defibrillation Electrode
Lead Placement Epi Patch
Lead Tip Location n/a
Pace/Sense Polarity n/a
More
Serial Number Prefix TAB, TAC, TAD
Lead Body Diameter S, M, L
Insulation Silicone
Pace/Sense Connector Style n/a
High Voltage Connector 6.5 mm
Steroid None
Hide this content

Data as of March 12, 2013