6721 Epicardial Patch
Defibrillation
6721 Epicardial Patch
ALL
Information

Device Survival Probability


6721_6921_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 111 mo
% Survival 96.5 95.0 92.7 91.9 90.0 85.1 83.7 83.7 83.7 83.7
#   330 301 256 209 176 132 95 66 56 50

System Longevity Study Results

Number of Leads Enrolled in Study 407
Cumulative Months of Follow-Up 23303
Number of Leads Active in Study 4
Information 47

Qualifying Complications

Conductor Fracture 21
Failure to Capture 8
Impedance Out of Range 4
Insulation (not further defined) 2
Oversensing 12
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

Information

US Acute Lead Observations

Cardiac Perforation 1
Conductor Fracture 2
Failure to Sense 1
Impedance Out of Range 3
Oversensing 1
Information

Distribution Data

US Market Release1994-03-31
CE Approval Date1993-01-01
Registered USA Implants2787
Estimated Active USA Implants1069
US Market Release1994-03-31
CE Approval Date1993-01-01
Estimated WW Distribution3970
Information

USA Returned Product Analysis

Conductor Fracture 12
Insulation Breach 1
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 15
Insulation Breach 1
Crimp/Weld/Bond 0
Other 0
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Suture
Lead Function Defibrillation Electrode
Lead Placement Epi Patch
Lead Tip Location n/a
Pace/Sense Polarity n/a
More
Serial Number Prefix TBH, TBG, TBB
Lead Body Diameter S, M, L
Insulation Silicone
Pace/Sense Connector Style n/a
High Voltage Connector DF-1
Steroid None
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050