5594 CapSure SP Novus
Pacing
5594 CapSure SP Novus
ALL
Information

Device Survival Probability


5594_SURV
    at 0 mo                  
% Survival 100.0                  
#   20                  

System Longevity Study Results

Number of Leads Enrolled in Study 20
Cumulative Months of Follow-Up 1515
Number of Leads Active in Study 10
Information 0

Qualifying Complications

Cardiac Perforation 0
Conductor Fracture 0
Electrical Abandonment 0
Extra Cardiac Stimulation 0
Failure to Capture 0
Failure to Sense 0
Impedance Out of Range 0
Insulation (ESC) 0
Insulation (MIO) 0
Insulation (not further defined) 0
Lead Dislodgement 0
Medical Judgement 0
Oversensing 0
Other 0
Unspecified Clinical Failure 0
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Lead Dislodgement 8
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release2001-06-25
CE Approval Date2001-03-23
Registered USA Implants15765
Estimated Active USA Implants10258
US Market Release2001-06-25
CE Approval Date2001-03-23
Estimated WW Distribution47860
Information

USA Returned Product Analysis

Conductor Fracture 5
Insulation Breach 7
Crimp/Weld/Bond 0
Other 1
Information

WW Returned Product Analysis

Conductor Fracture 8
Insulation Breach 7
Crimp/Weld/Bond 0
Other 2
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium - J
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LFD
Lead Body Diameter 6.0 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050