5592 CapSure SP Novus
Pacing
5592 CapSure SP Novus
ALL
Information

Device Survival Probability


5592_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 105 mo  
% Survival 99.7 99.3 99.3 98.9 98.9 98.9 98.9 98.9 98.9  
#   517 434 347 283 219 153 111 81 54  

System Longevity Study Results

Number of Leads Enrolled in Study 672
Cumulative Months of Follow-Up 31062
Number of Leads Active in Study 112
Information 5

Qualifying Complications

Failure to Capture 3
Lead Dislodgement 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Failure to Capture 4
Failure to Sense 2
Lead Dislodgement 24
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1998-06-03
CE Approval Date1997-09-25
Registered USA Implants34543
Estimated Active USA Implants18944
US Market Release1998-06-03
CE Approval Date1997-09-25
Estimated WW Distribution102111
Information

USA Returned Product Analysis

Conductor Fracture 4
Insulation Breach 4
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 5
Insulation Breach 4
Crimp/Weld/Bond 0
Other 0
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium - J
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LEU
Lead Body Diameter 6.0 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050