5568 CapSureFix
Pacing
5568 CapSureFix
ALL
Information

Device Survival Probability


5568_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr                    
% Survival 99.7 99.2 99.2 99.2 98.8 97.8 97.8 97.8 97.8 97.8 97.8                    
#   712 526 402 287 222 170 136 114 95 72 45                    

System Longevity Study Results

Number of Leads Enrolled in Study 1053
Cumulative Months of Follow-Up 40661
Number of Leads Active in Study 99
Information 11

Qualifying Complications

Conductor Fracture 1
Extra Cardiac Stimulation 1
Failure to Capture 5
Failure to Sense 2
Lead Dislodgement 1
Oversensing 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 7
Extra Cardiac Stimulation 2
Failure to Capture 19
Failure to Sense 2
Impedance Out of Range 2
Insulation Breach 1
Lead Dislodgement 30
Oversensing 2
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release1997-01-02
CE Approval Date1996-08-14
Registered USA Implants90933
Estimated Active USA Implants50511
US Market Release1997-01-02
CE Approval Date1996-08-14
Estimated WW Distribution143883
Information

USA Returned Product Analysis

Conductor Fracture 13
Insulation Breach 28
Crimp/Weld/Bond 0
Other 38
Information

WW Returned Product Analysis

Conductor Fracture 14
Insulation Breach 29
Crimp/Weld/Bond 0
Other 96
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Active Screw-in
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium - J
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LDN
Lead Body Diameter 7.2 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050