Device Survival Probability


5554_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 78 mo      
% Survival 100.0 99.2 98.7 98.2 98.2 98.2 98.2      
#   273 227 182 159 137 80 49      

System Longevity Study Results

Number of Leads Enrolled in Study 344
Cumulative Months of Follow-Up 15784
Number of Leads Active in Study 35
Information 4

Qualifying Complications

Failure to Capture 1
Impedance Out of Range 1
Lead Dislodgement 1
Oversensing 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Conductor Fracture 1
Failure to Capture 25
Failure to Sense 2
Lead Dislodgement 29
Unspecified Clinical Failure 3
Information

Distribution Data

US Market Release1998-06-03
CE Approval Date1997-06-05
Registered USA Implants62013
Estimated Active USA Implants29209
US Market Release1998-06-03
CE Approval Date1997-06-05
Estimated WW Distribution220963
Information

USA Returned Product Analysis

Conductor Fracture 9
Insulation Breach 18
Crimp/Weld/Bond 0
Other 1
Information

WW Returned Product Analysis

Conductor Fracture 36
Insulation Breach 25
Crimp/Weld/Bond 0
Other 3
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium - J
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LEJ
Lead Body Diameter 6.0 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050