5534 CapSure Z
Pacing
5534 CapSure Z
ALL
Information

Device Survival Probability


5534_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 78 mo      
% Survival 98.3 97.8 97.8 97.1 97.1 97.1 97.1      
#   207 182 155 108 72 56 51      

System Longevity Study Results

Number of Leads Enrolled in Study 264
Cumulative Months of Follow-Up 13009
Number of Leads Active in Study 6
Information 6

Qualifying Complications

Failure to Capture 5
Impedance Out of Range 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Extra Cardiac Stimulation 1
Failure to Capture 3
Failure to Sense 1
Lead Dislodgement 6
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release1996-02-09
CE Approval Date
Registered USA Implants25776
Estimated Active USA Implants6800
US Market Release1996-02-09
CE Approval Date
Estimated WW Distribution30752
Information

USA Returned Product Analysis

Conductor Fracture 3
Insulation Breach 4
Crimp/Weld/Bond 0
Other 4
Information

WW Returned Product Analysis

Conductor Fracture 6
Insulation Breach 6
Crimp/Weld/Bond 0
Other 7
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium - J
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LDG
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050