5524 CapSure SP
Pacing
5524 CapSure SP
ALL
Information

Device Survival Probability


5524_5524M_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr at 195 mo        
% Survival 99.7 99.7 99.5 99.3 99.2 99.2 98.9 98.4 98.1 97.8 97.6 97.6 97.6 97.0 96.2 96.2 96.2        
#   3438 2964 2541 2136 1781 1418 1120 862 691 556 422 311 233 170 102 53 49        

System Longevity Study Results

Number of Leads Enrolled in Study 4496
Cumulative Months of Follow-Up 255338
Number of Leads Active in Study 233
Information 41

Qualifying Complications

Conductor Fracture 1
Failure to Capture 23
Failure to Sense 4
Impedance Out of Range 1
Insulation (not further defined) 1
Lead Dislodgement 4
Oversensing 6
Other 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Information

Distribution Data

US Market Release1990-03-01
CE Approval Date
Registered USA Implants462
Estimated Active USA Implants103
US Market Release1990-03-01
CE Approval Date
Estimated WW Distribution517
Information

USA Returned Product Analysis

Conductor Fracture 0
Insulation Breach 0
Crimp/Weld/Bond 1
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 0
Insulation Breach 0
Crimp/Weld/Bond 1
Other 0
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium - J
Pace/Sense Polarity Bipolar
More
Serial Number Prefix XV
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style 3.2 mm
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050