5092 CapSure SP Novus
Pacing
5092 CapSure SP Novus
ALL
Information

Device Survival Probability


5092_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 117 mo
% Survival 99.5 99.4 99.2 99.0 99.0 98.6 98.6 98.6 98.6 98.6
#   812 650 518 407 316 240 179 123 79 52

System Longevity Study Results

Number of Leads Enrolled in Study 1172
Cumulative Months of Follow-Up 47860
Number of Leads Active in Study 130
Information 9

Qualifying Complications

Extra Cardiac Stimulation 1
Failure to Capture 2
Impedance Out of Range 1
Lead Dislodgement 5
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 4
Conductor Fracture 1
Extra Cardiac Stimulation 3
Failure to Capture 30
Failure to Sense 6
Insulation Breach 3
Lead Dislodgement 43
Oversensing 1
Unspecified Clinical Failure 9
Information

Distribution Data

US Market Release1998-06-03
CE Approval Date1997-09-25
Registered USA Implants132748
Estimated Active USA Implants60768
US Market Release1998-06-03
CE Approval Date1997-09-25
Estimated WW Distribution353308
Information

USA Returned Product Analysis

Conductor Fracture 10
Insulation Breach 38
Crimp/Weld/Bond 0
Other 5
Information

WW Returned Product Analysis

Conductor Fracture 10
Insulation Breach 40
Crimp/Weld/Bond 0
Other 16
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LET
Lead Body Diameter 6.0 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050