5076
CapSureFix Novus
VENTRICULAR PLACEMENT
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | at 117 mo | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.6 | 99.4 | 99.3 | 99.0 | 99.0 | 98.8 | 98.8 | 98.8 | 98.0 | 98.0 | ||||||||||||
| # | 1082 | 901 | 744 | 612 | 492 | 394 | 280 | 178 | 90 | 51 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 1539 |
|---|---|
| Cumulative Months of Follow-Up | 67711 |
| Number of Leads Active in Study | 276 |
ATRIAL PLACEMENT
Device Survival Probability
5076_ATR_SURV
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | at 123 mo | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.6 | 99.6 | 99.3 | 99.0 | 98.8 | 98.6 | 98.6 | 98.5 | 98.5 | 97.8 | 97.8 | |||||||||||
| # | 2084 | 1784 | 1527 | 1314 | 1112 | 902 | 624 | 391 | 215 | 82 | 53 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 2744 |
|---|---|
| Cumulative Months of Follow-Up | 139060 |
| Number of Leads Active in Study | 648 |
Performance Notes For This Model
US Acute Lead Observations
| Cardiac Perforation | 150 |
|---|---|
| Conductor Fracture | 11 |
| Extra Cardiac Stimulation | 13 |
| Failure to Capture | 227 |
| Failure to Sense | 35 |
| Impedance Out of Range | 12 |
| Insulation Breach | 8 |
| Lead Dislodgement | 470 |
| Oversensing | 27 |
| Unspecified Clinical Failure | 31 |
Product Characteristics
| Lead Type | Pacing |
|---|---|
| Type and/or Fixation | Active Screw-in |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Atrium or Right Ventricle |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | PJN |
|---|---|
| Lead Body Diameter | 6.0 FR |
| Insulation | Silicone |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Steroid |
Data as of March 12, 2013