5076 CapSureFix Novus
Pacing
5076 CapSureFix Novus
VENTRICULAR PLACEMENT
Information

Device Survival Probability


5076_VEN_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 117 mo                      
% Survival 99.6 99.4 99.3 99.0 99.0 98.8 98.8 98.8 98.0 98.0                      
#   1082 901 744 612 492 394 280 178 90 51                      

System Longevity Study Results

Number of Leads Enrolled in Study 1539
Cumulative Months of Follow-Up 67711
Number of Leads Active in Study 276
Information 12

Qualifying Complications

Cardiac Perforation 1
Conductor Fracture 1
Failure to Capture 5
Failure to Sense 1
Impedance Out of Range 2
Lead Dislodgement 2
ATRIAL PLACEMENT
Information

Device Survival Probability


5076_ATR_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 123 mo                    
% Survival 99.6 99.6 99.3 99.0 98.8 98.6 98.6 98.5 98.5 97.8 97.8                    
#   2084 1784 1527 1314 1112 902 624 391 215 82 53                    

System Longevity Study Results

Number of Leads Enrolled in Study 2744
Cumulative Months of Follow-Up 139060
Number of Leads Active in Study 648
Information 26

Qualifying Complications

Cardiac Perforation 1
Conductor Fracture 2
Extra Cardiac Stimulation 2
Failure to Capture 6
Failure to Sense 2
Impedance Out of Range 4
Insulation (not further defined) 1
Lead Dislodgement 5
Oversensing 2
Unspecified Clinical Failure 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 150
Conductor Fracture 11
Extra Cardiac Stimulation 13
Failure to Capture 227
Failure to Sense 35
Impedance Out of Range 12
Insulation Breach 8
Lead Dislodgement 470
Oversensing 27
Unspecified Clinical Failure 31
Information

Distribution Data

US Market Release2000-08-31
CE Approval Date1999-08-12
Registered USA Implants1511222
Estimated Active USA Implants956132
US Market Release2000-08-31
CE Approval Date1999-08-12
Estimated WW Distribution2614664
Information

USA Returned Product Analysis

Conductor Fracture 404
Insulation Breach 422
Crimp/Weld/Bond 0
Other 204
Information

WW Returned Product Analysis

Conductor Fracture 451
Insulation Breach 459
Crimp/Weld/Bond 0
Other 297
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Active Screw-in
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix PJN
Lead Body Diameter 6.0 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050