5071
Screw-in
ALL
Device Survival Probability
5071_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | at 90.0 mo | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | 95.2 | 92.1 | 90.4 | 90.4 | 89.0 | 87.4 | 86.3 | 84.8 | |||
# | 235 | 184 | 158 | 142 | 121 | 93 | 72 | 59 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 470 |
---|---|
Cumulative Months of Follow-Up | 17254 |
Number of Leads Active in Study | 64 |
US Acute Lead Observations
Cardiac Perforation | 2 |
---|---|
Extra Cardiac Stimulation | 6 |
Failure to Capture | 108 |
Failure to Sense | 4 |
Impedance Out of Range | 14 |
Lead Dislodgement | 2 |
Oversensing | 2 |
Unspecified Clinical Failure | 1 |
Product Characteristics
Lead Type | Epi/Myocardial |
---|---|
Type and/or Fixation | Fixed Screw |
Lead Function | Pacing/Sensing |
Lead Placement | Myocardial |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | Unipolar |
Serial Number Prefix | LAQ |
---|---|
Lead Body Diameter | 6.6 FR |
Insulation | Silicon |
Pace/Sense Connector Style | IS-1 UNI |
High Voltage Connector | n/a |
Steroid | None |
Data as of December 1, 2023