5071 Screw-in
Epi/Myocardial Pacing
5071 Screw-in
ALL
Information

Device Survival Probability

5071_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr        
% Survival 95.3 92.2 90.1 90.1 88.2 85.8        
#   231 173 137 107 73 52        

Product Surveillance Registry Results

Number of Leads Enrolled in Study 442
Cumulative Months of Follow-Up 12945
Number of Leads Active in Study 102
Information 31

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 1
Failure to Capture 20
Failure to Sense 2
Impedance Out of Range 1
Lead Dislodgement 1
Oversensing 2
Other 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Extra Cardiac Stimulation 6
Failure to Capture 78
Failure to Sense 3
Impedance Out of Range 6
Lead Dislodgement 1
Oversensing 1
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1992-12-03
CE Approval Date1993-01-01
Registered USA Implants53859
Estimated Active USA Implants16928
US Market Release1992-12-03
CE Approval Date1993-01-01
Estimated WW Distribution103490
Information

USA Returned Product Analysis

Conductor Fracture 24
Insulation Breach 2
Crimp/Weld/Bond 0
Other 1
Information

WW Returned Product Analysis

Conductor Fracture 25
Insulation Breach 3
Crimp/Weld/Bond 1
Other 5
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Fixed Screw
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LAQ
Lead Body Diameter 6.6 FR
Insulation Silicon
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid None
Hide this content

Data as of April 11, 2019

www.medtronic.com/productperformance 651050