5071 Screw-in
Epi/Myocardial Pacing
5071 Screw-in
ALL
Information

Device Survival Probability

5071_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr at 54.0 mo          
% Survival 96.7 93.6 90.3 90.3 90.3          
#   160 109 76 60 51          

Product Surveillance Registry Results

Number of Leads Enrolled in Study 375
Cumulative Months of Follow-Up 8274
Number of Leads Active in Study 88
Information 20

Qualifying Complications

Conductor Fracture 1
Failure to Capture 15
Impedance Out of Range 1
Lead Dislodgement 1
Oversensing 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Extra Cardiac Stimulation 5
Failure to Capture 48
Failure to Sense 2
Impedance Out of Range 3
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1992-12-03
CE Approval Date1993-01-01
Registered USA Implants49214
Estimated Active USA Implants15137
US Market Release1992-12-03
CE Approval Date1993-01-01
Estimated WW Distribution92437
Information

USA Returned Product Analysis

Conductor Fracture 17
Insulation Breach 2
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 18
Insulation Breach 3
Crimp/Weld/Bond 1
Other 9
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Fixed Screw In
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LAQ
Lead Body Diameter 6.6 FR
Insulation Silicon
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid None
Hide this content

Data as of November 2, 2015

www.medtronic.com/productperformance 376051