Device Survival Probability

5071_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr at 66.0 mo        
% Survival 95.5 92.0 89.7 89.7 88.5 88.5        
#   214 158 125 91 63 50        

Product Surveillance Registry Results

Number of Leads Enrolled in Study 430
Cumulative Months of Follow-Up 11626
Number of Leads Active in Study 107
Information 28

Qualifying Complications

Conductor Fracture 3
Failure to Capture 19
Failure to Sense 2
Impedance Out of Range 1
Lead Dislodgement 1
Oversensing 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Extra Cardiac Stimulation 6
Failure to Capture 63
Failure to Sense 3
Impedance Out of Range 6
Oversensing 1
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1992-12-03
CE Approval Date1993-01-01
Registered USA Implants52663
Estimated Active USA Implants16397
US Market Release1992-12-03
CE Approval Date1993-01-01
Estimated WW Distribution100552
Information

USA Returned Product Analysis

Conductor Fracture 24
Insulation Breach 2
Crimp/Weld/Bond 0
Other 1
Information

WW Returned Product Analysis

Conductor Fracture 25
Insulation Breach 3
Crimp/Weld/Bond 1
Other 10
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Fixed Screw
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LAQ
Lead Body Diameter 6.6 FR
Insulation Silicon
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid None
Hide this content

Data as of January 31, 2018