5071 Screw-in
Epi/Myocardial Pacing
5071 Screw-in
ALL
Information

Device Survival Probability

5071_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr at 54.0 mo          
% Survival 95.1 91.5 88.3 88.3 88.3          
#   173 118 78 63 54          

Product Surveillance Registry Results

Number of Leads Enrolled in Study 386
Cumulative Months of Follow-Up 8886
Number of Leads Active in Study 93
Information 25

Qualifying Complications

Conductor Fracture 1
Failure to Capture 17
Failure to Sense 2
Impedance Out of Range 1
Lead Dislodgement 1
Oversensing 2
Other 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Extra Cardiac Stimulation 5
Failure to Capture 50
Failure to Sense 3
Impedance Out of Range 3
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1992-12-03
CE Approval Date1993-01-01
Registered USA Implants49964
Estimated Active USA Implants15301
US Market Release1992-12-03
CE Approval Date1993-01-01
Estimated WW Distribution94024
Information

USA Returned Product Analysis

Conductor Fracture 18
Insulation Breach 3
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 19
Insulation Breach 4
Crimp/Weld/Bond 1
Other 9
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Fixed Screw In
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LAQ
Lead Body Diameter 6.6 FR
Insulation Silicon
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid None
Hide this content

Data as of April 14, 2016

www.medtronic.com/productperformance 426050