5068 CapSureFix
Pacing
5068 CapSureFix
VENTRICULAR PLACEMENT
Information

Device Survival Probability


5068_VEN_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 126 mo                    
% Survival 99.9 99.7 99.5 99.2 99.2 99.2 98.5 98.5 98.5 98.5 97.6                    
#   700 499 363 268 227 189 152 128 102 70 47                    

System Longevity Study Results

Number of Leads Enrolled in Study 1362
Cumulative Months of Follow-Up 40437
Number of Leads Active in Study 65
Information 5

Qualifying Complications

Conductor Fracture 1
Failure to Capture 2
Insulation (not further defined) 1
Lead Dislodgement 1
ATRIAL PLACEMENT
Information

Device Survival Probability


5068_ATR_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 123 mo                    
% Survival 99.6 99.6 99.3 99.3 99.3 99.3 99.3 97.6 97.6 97.6 97.6                    
#   553 425 322 254 211 160 131 104 78 62 55                    

System Longevity Study Results

Number of Leads Enrolled in Study 968
Cumulative Months of Follow-Up 34032
Number of Leads Active in Study 31
Information 6

Qualifying Complications

Failure to Capture 2
Impedance Out of Range 1
Insulation (not further defined) 1
Lead Dislodgement 1
Oversensing 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 16
Conductor Fracture 4
Failure to Capture 31
Failure to Sense 5
Impedance Out of Range 1
Insulation Breach 1
Lead Dislodgement 20
Oversensing 1
Unspecified Clinical Failure 6
Information

Distribution Data

US Market Release1997-01-02
CE Approval Date
Registered USA Implants102363
Estimated Active USA Implants29266
US Market Release1997-01-02
CE Approval Date
Estimated WW Distribution135015
Information

USA Returned Product Analysis

Conductor Fracture 40
Insulation Breach 56
Crimp/Weld/Bond 2
Other 83
Information

WW Returned Product Analysis

Conductor Fracture 41
Insulation Breach 62
Crimp/Weld/Bond 2
Other 198
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Active Screw-in
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LDJ
Lead Body Diameter 7.2 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050