5068
CapSureFix
VENTRICULAR PLACEMENT
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | at 126 mo | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.9 | 99.7 | 99.5 | 99.2 | 99.2 | 99.2 | 98.5 | 98.5 | 98.5 | 98.5 | 97.6 | |||||||||||
| # | 700 | 499 | 363 | 268 | 227 | 189 | 152 | 128 | 102 | 70 | 47 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 1362 |
|---|---|
| Cumulative Months of Follow-Up | 40437 |
| Number of Leads Active in Study | 65 |
ATRIAL PLACEMENT
Device Survival Probability
5068_ATR_SURV
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | at 123 mo | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.6 | 99.6 | 99.3 | 99.3 | 99.3 | 99.3 | 99.3 | 97.6 | 97.6 | 97.6 | 97.6 | |||||||||||
| # | 553 | 425 | 322 | 254 | 211 | 160 | 131 | 104 | 78 | 62 | 55 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 968 |
|---|---|
| Cumulative Months of Follow-Up | 34032 |
| Number of Leads Active in Study | 31 |
Performance Notes For This Model
US Acute Lead Observations
| Cardiac Perforation | 16 |
|---|---|
| Conductor Fracture | 4 |
| Failure to Capture | 31 |
| Failure to Sense | 5 |
| Impedance Out of Range | 1 |
| Insulation Breach | 1 |
| Lead Dislodgement | 20 |
| Oversensing | 1 |
| Unspecified Clinical Failure | 6 |
Product Characteristics
| Lead Type | Pacing |
|---|---|
| Type and/or Fixation | Active Screw-in |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Atrium or Right Ventricle |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | LDJ |
|---|---|
| Lead Body Diameter | 7.2 FR |
| Insulation | Silicone |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Steroid |
Data as of March 12, 2013