5054
CapSure Z Novus
VENTRICULAR PLACEMENT
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | at 117 mo | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.5 | 99.4 | 99.4 | 99.1 | 99.1 | 97.9 | 97.9 | 97.9 | 97.9 | 97.9 | ||||||||||||
| # | 656 | 513 | 387 | 318 | 274 | 202 | 157 | 116 | 76 | 49 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 967 |
|---|---|
| Cumulative Months of Follow-Up | 39207 |
| Number of Leads Active in Study | 92 |
ATRIAL PLACEMENT
Device Survival Probability
5054_ATR_SURV
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | ||||||||||||
| # | 408 | 385 | 357 | 322 | 284 | 246 | 208 | 155 | 93 | 50 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 424 |
|---|---|
| Cumulative Months of Follow-Up | 33362 |
| Number of Leads Active in Study | 156 |
Performance Notes For This Model
US Acute Lead Observations
| Cardiac Perforation | 1 |
|---|---|
| Conductor Fracture | 1 |
| Failure to Capture | 22 |
| Impedance Out of Range | 2 |
| Insulation Breach | 1 |
| Lead Dislodgement | 22 |
| Unspecified Clinical Failure | 9 |
Product Characteristics
| Lead Type | Pacing |
|---|---|
| Type and/or Fixation | Tines |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Atrium or Right Ventricle |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | LEH |
|---|---|
| Lead Body Diameter | 6.0 FR |
| Insulation | Silicone |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Steroid |
Data as of March 12, 2013