5054 CapSure Z Novus
Pacing
5054 CapSure Z Novus
VENTRICULAR PLACEMENT
Information

Device Survival Probability

5054_VEN_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr                  
% Survival 99.3 99.1 98.8 98.4 98.4 97.0 97.0 96.4 96.4 96.4 96.4 96.4                  
#   474 391 304 264 230 192 167 144 113 93 72 54                  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 991
Cumulative Months of Follow-Up 35502
Number of Leads Active in Study 18
Information 13

Qualifying Complications

Failure to Capture 7
Failure to Sense 2
Impedance Out of Range 1
Lead Dislodgement 1
Other 2
ATRIAL PLACEMENT
Information

Device Survival Probability

5054_ATR_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr at 198.0 mo        
% Survival 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5 99.5        
#   411 391 358 322 289 252 219 185 152 128 107 92 74 64 58 51 50        

Product Surveillance Registry Results

Number of Leads Enrolled in Study 425
Cumulative Months of Follow-Up 41984
Number of Leads Active in Study 28
Information 3

Qualifying Complications

Failure to Capture 2
Impedance Out of Range 0
Lead Dislodgement 1
Other 0
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 2
Failure to Capture 23
Impedance Out of Range 4
Insulation Breach 1
Lead Dislodgement 30
Unspecified Clinical Failure 9
Information

Distribution Data

US Market Release1998-06-03
CE Approval Date1997-06-05
Registered USA Implants100058
Estimated Active USA Implants18503
US Market Release1998-06-03
CE Approval Date1997-06-05
Estimated WW Distribution500745
Information

USA Returned Product Analysis

Conductor Fracture 16
Insulation Breach 46
Crimp/Weld/Bond 1
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 22
Insulation Breach 52
Crimp/Weld/Bond 1
Other 5
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LEH
Lead Body Diameter 6.0 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of December 1, 2023

www.medtronic.com/productperformance 951051