Device Survival Probability


5038_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 81 mo      
% Survival 99.8 99.8 99.8 98.8 98.2 98.2 95.2      
#   420 308 220 176 135 83 52      

System Longevity Study Results

Number of Leads Enrolled in Study 558
Cumulative Months of Follow-Up 20718
Number of Leads Active in Study 51
Information 6

Qualifying Complications

Conductor Fracture 3
Failure to Capture 1
Failure to Sense 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Extra Cardiac Stimulation 1
Failure to Capture 1
Failure to Sense 1
Lead Dislodgement 2
Information

Distribution Data

US Market Release1998-09-10
CE Approval Date1997-04-15
Registered USA Implants9093
Estimated Active USA Implants3527
US Market Release1998-09-10
CE Approval Date1997-04-15
Estimated WW Distribution111854
Information

USA Returned Product Analysis

Conductor Fracture 4
Insulation Breach 1
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 14
Insulation Breach 10
Crimp/Weld/Bond 0
Other 11
Information

Product Characteristics

Lead Type VDD Single Pass
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LEG, LEE, LEF
Lead Body Diameter 8.1 FR
Insulation Silicone
Pace/Sense Connector Style Two IS-1 Bi
High Voltage Connector n/a
Steroid Yes
Hide this content