5034
CapSure Z
VENTRICULAR PLACEMENT
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.7 | 99.4 | 99.1 | 98.8 | 98.8 | 98.8 | 98.1 | 96.6 | ||||||||||||||
| # | 886 | 671 | 507 | 340 | 211 | 127 | 77 | 50 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 1209 |
|---|---|
| Cumulative Months of Follow-Up | 44496 |
| Number of Leads Active in Study | 12 |
ATRIAL PLACEMENT
Device Survival Probability
5034_ATR_SURV
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | at 159 mo | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 98.8 | 98.8 | 98.8 | 98.8 | 98.8 | 98.8 | 97.5 | 97.5 | ||||||||
| # | 383 | 382 | 379 | 373 | 355 | 330 | 294 | 250 | 207 | 183 | 147 | 102 | 61 | 50 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 386 |
|---|---|
| Cumulative Months of Follow-Up | 44637 |
| Number of Leads Active in Study | 130 |
Performance Notes For This Model
US Acute Lead Observations
| Cardiac Perforation | 2 |
|---|---|
| Conductor Fracture | 2 |
| Failure to Capture | 28 |
| Failure to Sense | 3 |
| Insulation Breach | 3 |
| Lead Dislodgement | 14 |
| Unspecified Clinical Failure | 12 |
Product Characteristics
| Lead Type | Pacing |
|---|---|
| Type and/or Fixation | Tines |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Atrium or Right Ventricle |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | LDF |
|---|---|
| Lead Body Diameter | 7.8 FR |
| Insulation | Silicone |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Steroid |
Data as of March 12, 2013