5034 CapSure Z
Pacing
5034 CapSure Z
VENTRICULAR PLACEMENT
Information

Device Survival Probability


5034_VEN_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr                          
% Survival 99.7 99.4 99.1 98.8 98.8 98.8 98.1 96.6                          
#   886 671 507 340 211 127 77 50                          

System Longevity Study Results

Number of Leads Enrolled in Study 1209
Cumulative Months of Follow-Up 44496
Number of Leads Active in Study 12
Information 11

Qualifying Complications

Conductor Fracture 1
Failure to Capture 7
Failure to Sense 2
Lead Dislodgement 1
ATRIAL PLACEMENT
Information

Device Survival Probability


5034_ATR_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr at 159 mo              
% Survival 99.5 99.5 99.5 99.5 99.5 99.5 98.8 98.8 98.8 98.8 98.8 98.8 97.5 97.5              
#   383 382 379 373 355 330 294 250 207 183 147 102 61 50              

System Longevity Study Results

Number of Leads Enrolled in Study 386
Cumulative Months of Follow-Up 44637
Number of Leads Active in Study 130
Information 5

Qualifying Complications

Conductor Fracture 1
Failure to Capture 2
Failure to Sense 1
Impedance Out of Range 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 2
Failure to Capture 28
Failure to Sense 3
Insulation Breach 3
Lead Dislodgement 14
Unspecified Clinical Failure 12
Information

Distribution Data

US Market Release1996-02-09
CE Approval Date
Registered USA Implants55236
Estimated Active USA Implants12331
US Market Release1996-02-09
CE Approval Date
Estimated WW Distribution69347
Information

USA Returned Product Analysis

Conductor Fracture 12
Insulation Breach 14
Crimp/Weld/Bond 0
Other 7
Information

WW Returned Product Analysis

Conductor Fracture 17
Insulation Breach 18
Crimp/Weld/Bond 0
Other 16
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LDF
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050