5033 CapSure Z
Pacing
5033 CapSure Z
ALL
Information

Device Survival Probability


5033_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr at 177 mo            
% Survival 99.7 99.6 99.1 99.0 98.8 98.3 97.7 96.9 96.0 96.0 95.6 95.6 94.9 93.9 93.9            
#   1415 1143 944 795 658 527 416 347 290 240 188 153 126 78 45            

System Longevity Study Results

Number of Leads Enrolled in Study 1899
Cumulative Months of Follow-Up 100842
Number of Leads Active in Study 148
Information 29

Qualifying Complications

Cardiac Perforation 1
Conductor Fracture 8
Failure to Capture 13
Impedance Out of Range 4
Insulation (not further defined) 1
Lead Dislodgement 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1996-02-09
CE Approval Date
Registered USA Implants2336
Estimated Active USA Implants479
US Market Release1996-02-09
CE Approval Date
Estimated WW Distribution9278
Information

USA Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 3
Information

WW Returned Product Analysis

Conductor Fracture 2
Insulation Breach 1
Crimp/Weld/Bond 0
Other 5
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LDK
Lead Body Diameter 5.1 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013