5032 CapSure VDD
VDD Single Pass Pacing
5032 CapSure VDD
ALL
Information

Device Survival Probability


5032_SURV
    at 0 mo                  
% Survival 100.0                  
#   37                  

System Longevity Study Results

Number of Leads Enrolled in Study 38
Cumulative Months of Follow-Up 1683
Number of Leads Active in Study 0
Information 1

Qualifying Complications

Failure to Sense 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Failure to Capture 1
Failure to Sense 1
Lead Dislodgement 1
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1996-03-22
CE Approval Date
Registered USA Implants5307
Estimated Active USA Implants1209
US Market Release1996-03-22
CE Approval Date
Estimated WW Distribution9005
Information

USA Returned Product Analysis

Conductor Fracture 7
Insulation Breach 6
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 10
Insulation Breach 8
Crimp/Weld/Bond 1
Other 0
Information

Product Characteristics

Lead Type VDD Single Pass
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Quadripolar
More
Serial Number Prefix LCN, LCM, LCL
Lead Body Diameter 8.1 FR
Insulation Silicone
Pace/Sense Connector Style Two IS-1 Bi
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050