5024M CapSure SP
Pacing
5024M CapSure SP
ALL
Information

Device Survival Probability


5024_5024M_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr at 207 mo      
% Survival 99.6 99.6 99.5 99.4 99.3 99.3 99.1 99.0 98.9 98.8 98.5 98.5 98.3 97.9 97.1 97.1 95.6 95.6      
#   6128 5277 4477 3747 3113 2460 1927 1457 1103 837 636 466 348 250 163 102 56 48      

System Longevity Study Results

Number of Leads Enrolled in Study 8153
Cumulative Months of Follow-Up 444024
Number of Leads Active in Study 262
Information 57

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 2
Failure to Capture 27
Failure to Sense 2
Impedance Out of Range 3
Insulation (ESC) 1
Insulation (not further defined) 5
Lead Dislodgement 6
Oversensing 5
Other 2
Unspecified Clinical Failure 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 9
Conductor Fracture 8
Extra Cardiac Stimulation 3
Failure to Capture 49
Failure to Sense 9
Impedance Out of Range 3
Insulation Breach 6
Lead Dislodgement 30
Oversensing 1
Unspecified Clinical Failure 16
Information

Distribution Data

US Market Release1991-02-01
CE Approval Date
Registered USA Implants197022
Estimated Active USA Implants42313
US Market Release1991-02-01
CE Approval Date
Estimated WW Distribution231668
Information

USA Returned Product Analysis

Conductor Fracture 53
Insulation Breach 55
Crimp/Weld/Bond 0
Other 17
Information

WW Returned Product Analysis

Conductor Fracture 57
Insulation Breach 58
Crimp/Weld/Bond 7
Other 51
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LAT
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Steroid
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050