4968 CapSure EPI
Epi/Myocardial Pacing
4968 CapSure EPI
ALL
Information

Device Survival Probability


4968_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 117 mo
% Survival 98.6 96.7 95.1 93.0 92.1 91.2 89.2 89.2 86.4 84.7
#   618 525 428 342 266 195 142 99 66 47

System Longevity Study Results

Number of Leads Enrolled in Study 816
Cumulative Months of Follow-Up 39078
Number of Leads Active in Study 334
Information 61

Qualifying Complications

Conductor Fracture 12
Extra Cardiac Stimulation 2
Failure to Capture 20
Failure to Sense 4
Impedance Out of Range 5
Insulation (not further defined) 3
Oversensing 8
Other 7
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Failure to Capture 11
Impedance Out of Range 2
Insulation Breach 1
Lead Dislodgement 3
Oversensing 2
Information

Distribution Data

US Market Release1999-09-16
CE Approval Date1998-04-21
Registered USA Implants27775
Estimated Active USA Implants17196
US Market Release1999-09-16
CE Approval Date1998-04-21
Estimated WW Distribution87582
Information

USA Returned Product Analysis

Conductor Fracture 33
Insulation Breach 12
Crimp/Weld/Bond 0
Other 2
Information

WW Returned Product Analysis

Conductor Fracture 40
Insulation Breach 16
Crimp/Weld/Bond 0
Other 3
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Suture
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Myocardial
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LEN
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050