4965 CapSure Epi
Epi/Myocardial Pacing
4965 CapSure Epi
ALL
Information

Device Survival Probability

4965_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr at 54.0 mo          
% Survival 98.6 95.8 94.8 86.4 81.7          
#   132 113 93 68 50          

Product Surveillance Registry Results

Number of Leads Enrolled in Study 234
Cumulative Months of Follow-Up 7335
Number of Leads Active in Study 6
Information 17

Qualifying Complications

Conductor Fracture 10
Failure to Capture 3
Failure to Sense 1
Impedance Out of Range 0
Insulation (not further defined) 1
Oversensing 2
Other 0
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Conductor Fracture 1
Failure to Capture 10
Failure to Sense 5
Impedance Out of Range 16
Oversensing 1
Unspecified Clinical Failure 3
Information

Distribution Data

US Market Release1996-09-06
CE Approval Date1993-01-01
Registered USA Implants23375
Estimated Active USA Implants8278
US Market Release1996-09-06
CE Approval Date1993-01-01
Estimated WW Distribution57101
Information

USA Returned Product Analysis

Conductor Fracture 283
Insulation Breach 62
Crimp/Weld/Bond 1
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 341
Insulation Breach 70
Crimp/Weld/Bond 1
Other 0
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Suture
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LBT
Lead Body Diameter 4.5 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of April 10, 2020

www.medtronic.com/productperformance 751050