4965 CapSure Epi
Epi/Myocardial Pacing
4965 CapSure Epi
ALL
Information

Device Survival Probability

4965_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr            
% Survival 98.6 96.7 95.7 89.4            
#   131 111 91 67            

Product Surveillance Registry Results

Number of Leads Enrolled in Study 232
Cumulative Months of Follow-Up 7158
Number of Leads Active in Study 7
Information 14

Qualifying Complications

Conductor Fracture 7
Failure to Capture 3
Failure to Sense 1
Insulation (not further defined) 1
Oversensing 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Conductor Fracture 1
Failure to Capture 6
Failure to Sense 5
Impedance Out of Range 9
Oversensing 1
Unspecified Clinical Failure 3
Information

Distribution Data

US Market Release1996-09-06
CE Approval Date1993-01-01
Registered USA Implants22893
Estimated Active USA Implants8628
US Market Release1996-09-06
CE Approval Date1993-01-01
Estimated WW Distribution53962
Information

USA Returned Product Analysis

Conductor Fracture 247
Insulation Breach 50
Crimp/Weld/Bond 1
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 299
Insulation Breach 56
Crimp/Weld/Bond 1
Other 3
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Suture
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LBT
Lead Body Diameter 4.5 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of July 31, 2017

www.medtronic.com/productperformance 501050