4965 CapSure Epi
Epi/Myocardial Pacing
4965 CapSure Epi
ALL
Information

Device Survival Probability


4965_SURV
    1 yr 2 yr 3 yr 4 yr at 51 mo          
% Survival 99.0 98.2 97.2 92.7 91.3          
#   140 117 92 63 55          

System Longevity Study Results

Number of Leads Enrolled in Study 219
Cumulative Months of Follow-Up 7617
Number of Leads Active in Study 25
Information 12

Qualifying Complications

Conductor Fracture 5
Failure to Capture 3
Failure to Sense 1
Insulation (not further defined) 1
Oversensing 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Conductor Fracture 1
Failure to Capture 4
Failure to Sense 4
Impedance Out of Range 3
Oversensing 1
Unspecified Clinical Failure 3
Information

Distribution Data

US Market Release1996-09-06
CE Approval Date1993-01-01
Registered USA Implants20905
Estimated Active USA Implants9427
US Market Release1996-09-06
CE Approval Date1993-01-01
Estimated WW Distribution44904
Information

USA Returned Product Analysis

Conductor Fracture 163
Insulation Breach 34
Crimp/Weld/Bond 1
Other 2
Information

WW Returned Product Analysis

Conductor Fracture 203
Insulation Breach 38
Crimp/Weld/Bond 1
Other 5
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Suture
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Myocardial
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LBT
Lead Body Diameter 4.5 FR
Insulation Silicine
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050