4598
Attain Performa S
ALL
Device Survival Probability
4598_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | at 90.0 mo | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.2 | 99.0 | 98.9 | 98.6 | 98.5 | 98.5 | 98.5 | 97.0 | |||
# | 1126 | 950 | 782 | 602 | 383 | 220 | 105 | 55 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1364 |
---|---|
Cumulative Months of Follow-Up | 57890 |
Number of Leads Active in Study | 560 |
US Acute Lead Observations
Cardiac Perforation | 11 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 130 |
Failure to Capture | 92 |
Impedance Out of Range | 33 |
Lead Dislodgement | 82 |
Oversensing | 1 |
Product Characteristics
Lead Type | Left-Heart |
---|---|
Type and/or Fixation | S-shape |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Left Ventricular Cardiac Vein |
Pace/Sense Polarity | Quadripolar |
Serial Number Prefix | QUC |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) SI-polyimide (SI-PI)(inner) |
Pace/Sense Connector Style | IS4-LLLL |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of December 1, 2023