4592 CapSure SP Novus
Pacing
4592 CapSure SP Novus
ALL
Information

Device Survival Probability


4592_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr  
% Survival 98.2 98.2 98.2 98.2 97.4 97.4 97.4 96.0 96.0  
#   193 171 153 130 112 101 81 66 48  

System Longevity Study Results

Number of Leads Enrolled in Study 283
Cumulative Months of Follow-Up 14710
Number of Leads Active in Study 61
Information 6

Qualifying Complications

Failure to Capture 3
Failure to Sense 1
Lead Dislodgement 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Failure to Capture 8
Failure to Sense 1
Insulation Breach 1
Lead Dislodgement 26
Oversensing 1
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release1998-10-05
CE Approval Date1998-04-15
Registered USA Implants85431
Estimated Active USA Implants40834
US Market Release1998-10-05
CE Approval Date1998-04-15
Estimated WW Distribution180526
Information

USA Returned Product Analysis

Conductor Fracture 7
Insulation Breach 17
Crimp/Weld/Bond 0
Other 1
Information

WW Returned Product Analysis

Conductor Fracture 7
Insulation Breach 17
Crimp/Weld/Bond 0
Other 4
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation J-shape, tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium - J
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LER
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050