4574 CapSure Sense
Pacing
4574 CapSure Sense
ALL
Information

Device Survival Probability

4574_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 78.0 mo      
% Survival 99.2 99.2 99.2 98.9 98.6 98.6 98.6      
#   847 627 465 339 224 114 63      

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1106
Cumulative Months of Follow-Up 38371
Number of Leads Active in Study 607
Information 9

Qualifying Complications

Conductor Fracture 2
Failure to Capture 1
Lead Dislodgement 6
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Conductor Fracture 1
Extra Cardiac Stimulation 1
Failure to Capture 56
Failure to Sense 21
Impedance Out of Range 3
Lead Dislodgement 136
Oversensing 1
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release2002-06-23
CE Approval Date2002-02-01
Registered USA Implants85722
Estimated Active USA Implants55006
US Market Release2002-06-23
CE Approval Date2002-02-01
Estimated WW Distribution619656
Information

USA Returned Product Analysis

Conductor Fracture 10
Insulation Breach 13
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 13
Insulation Breach 13
Crimp/Weld/Bond 0
Other 2
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation J-shape, tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium
Pace/Sense Polarity Bipolar
More
Serial Number Prefix BBE
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of October 12, 2018

www.medtronic.com/productperformance 626050