4568 CapSureFix
Pacing
4568 CapSureFix
ALL
Information

Device Survival Probability


4568_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr  
% Survival 96.8 96.4 95.3 94.7 94.0 94.0 92.8 92.1 90.5  
#   495 431 350 292 242 187 134 80 51  

System Longevity Study Results

Number of Leads Enrolled in Study 656
Cumulative Months of Follow-Up 32177
Number of Leads Active in Study 133
Information 35

Qualifying Complications

Failure to Capture 18
Failure to Sense 4
Impedance Out of Range 3
Lead Dislodgement 9
Medical Judgement 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 3
Conductor Fracture 1
Failure to Capture 6
Failure to Sense 1
Impedance Out of Range 2
Lead Dislodgement 4
Oversensing 1
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1997-01-02
CE Approval Date
Registered USA Implants69137
Estimated Active USA Implants21496
US Market Release1997-01-02
CE Approval Date
Estimated WW Distribution92146
Information

USA Returned Product Analysis

Conductor Fracture 3
Insulation Breach 81
Crimp/Weld/Bond 0
Other 53
Information

WW Returned Product Analysis

Conductor Fracture 3
Insulation Breach 83
Crimp/Weld/Bond 0
Other 131
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation J-shape, screw in
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium - J
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LDD
Lead Body Diameter 7.5 FR
Insulation Polyurethane
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050