4396 Attain Ability Straight
Left-Heart Pacing
4396 Attain Ability Straight
ALL
Information

Device Survival Probability


4396_SURV
    at 9 mo                  
% Survival 100.0                  
#   45                  

System Longevity Study Results

Number of Leads Enrolled in Study 178
Cumulative Months of Follow-Up 1517
Number of Leads Active in Study 161
Information 0

Qualifying Complications

Cardiac Perforation 0
Conductor Fracture 0
Electrical Abandonment 0
Extra Cardiac Stimulation 0
Failure to Capture 0
Failure to Sense 0
Impedance Out of Range 0
Insulation (ESC) 0
Insulation (MIO) 0
Insulation (not further defined) 0
Lead Dislodgement 0
Medical Judgement 0
Oversensing 0
Other 0
Unspecified Clinical Failure 0
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Conductor Fracture 1
Extra Cardiac Stimulation 5
Failure to Capture 2
Lead Dislodgement 13
Information

Distribution Data

US Market Release2011-03-31
CE Approval Date2009-12-18
Registered USA Implants2803
Estimated Active USA Implants2602
US Market Release2011-03-31
CE Approval Date2009-12-18
Estimated WW Distribution7646
Information

USA Returned Product Analysis

Conductor Fracture
Insulation Breach
Crimp/Weld/Bond
Other
Information

WW Returned Product Analysis

Conductor Fracture
Insulation Breach
Crimp/Weld/Bond
Other
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Cardiac vein
Pace/Sense Polarity Dual Electrodes
More
Serial Number Prefix RAE
Lead Body Diameter 4.0 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050