Device Survival Probability

4298_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr at 54.0 mo          
% Survival 99.3 98.9 98.8 98.8 98.8          
#   1266 875 513 158 54          

Product Surveillance Registry Results

Number of Leads Enrolled in Study 2031
Cumulative Months of Follow-Up 44080
Number of Leads Active in Study 1473
Information 16

Qualifying Complications

Extra Cardiac Stimulation 4
Failure to Capture 0
Impedance Out of Range 0
Lead Dislodgement 11
Other 1
Information

US Acute Lead Observations

Cardiac Perforation 5
Conductor Fracture 1
Extra Cardiac Stimulation 170
Failure to Capture 93
Failure to Sense 1
Impedance Out of Range 23
Lead Dislodgement 148
Information

Distribution Data

US Market Release2014-08-01
CE Approval Date2013-01-01
Registered USA Implants75801
Estimated Active USA Implants70622
US Market Release2014-08-01
CE Approval Date2013-01-01
Estimated WW Distribution149230
Information

USA Returned Product Analysis

Conductor Fracture 3
Insulation Breach 0
Crimp/Weld/Bond 0
Other 15
Information

WW Returned Product Analysis

Conductor Fracture 5
Insulation Breach 0
Crimp/Weld/Bond 0
Other 17
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Bipolar
More
Serial Number Prefix QUA
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-4
High Voltage Connector n/a
Steroid Yes