4296 Attain Ability Plus
Left-Heart Pacing
4296 Attain Ability Plus
ALL
Information

Device Survival Probability


4296_SURV
    at 9 mo                  
% Survival 99.5                  
#   87                  

System Longevity Study Results

Number of Leads Enrolled in Study 672
Cumulative Months of Follow-Up 2876
Number of Leads Active in Study 618
Information 2

Qualifying Complications

Lead Dislodgement 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Extra Cardiac Stimulation 8
Failure to Capture 5
Insulation Breach 1
Lead Dislodgement 35
Information

Distribution Data

US Market Release2011-04-01
CE Approval Date2009-12-18
Registered USA Implants12038
Estimated Active USA Implants11344
US Market Release2011-04-01
CE Approval Date2009-12-18
Estimated WW Distribution27577
Information

USA Returned Product Analysis

Conductor Fracture
Insulation Breach
Crimp/Weld/Bond
Other
Information

WW Returned Product Analysis

Conductor Fracture 3
Insulation Breach 0
Crimp/Weld/Bond 0
Other 0
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Distal Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Cardiac vein
Pace/Sense Polarity Dual Electrodes
More
Serial Number Prefix RRA
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050