4296 Attain Ability Plus
Left-Heart Pacing
4296 Attain Ability Plus
ALL
Information

Device Survival Probability

4296_SURV
Loading...











    1 yr 2 yr 3 yr at 42.0 mo            
% Survival 98.8 98.0 98.0 98.0            
#   1029 587 231 96            

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1402
Cumulative Months of Follow-Up 30221
Number of Leads Active in Study 859
Information 22

Qualifying Complications

Extra Cardiac Stimulation 6
Failure to Capture 4
Lead Dislodgement 11
Other 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Extra Cardiac Stimulation 53
Failure to Capture 21
Impedance Out of Range 8
Insulation Breach 4
Lead Dislodgement 107
Information

Distribution Data

US Market Release2011-04-01
CE Approval Date2009-12-18
Registered USA Implants31226
Estimated Active USA Implants26671
US Market Release2011-04-01
CE Approval Date2009-12-18
Estimated WW Distribution89039
Information

USA Returned Product Analysis

Conductor Fracture 2
Insulation Breach 0
Crimp/Weld/Bond 2
Other 3
Information

WW Returned Product Analysis

Conductor Fracture 7
Insulation Breach 0
Crimp/Weld/Bond 2
Other 3
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Distal Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Dual Electrodes
More
Serial Number Prefix RRA
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of April 14, 2016