4196
Attain Ability
ALL
System Longevity Study Results
| Number of Leads Enrolled in Study | 1876 |
|---|---|
| Cumulative Months of Follow-Up | 34965 |
| Number of Leads Active in Study | 1300 |
Performance Notes For This Model
US Acute Lead Observations
| Conductor Fracture | 1 |
|---|---|
| Extra Cardiac Stimulation | 56 |
| Failure to Capture | 34 |
| Impedance Out of Range | 5 |
| Insulation Breach | 1 |
| Lead Dislodgement | 117 |
| Oversensing | 1 |
| Unspecified Clinical Failure | 3 |
Product Characteristics
| Lead Type | Left-Heart |
|---|---|
| Type and/or Fixation | Double Curve |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Cardiac vein |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | PVI |
|---|---|
| Lead Body Diameter | 4.0 FR |
| Insulation | Polyurethane (outer); SI-polyimide (inner) |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Yes |
Data as of March 12, 2013