4196
Attain Ability
ALL

Device Survival Probability
4196_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
98.0 | 97.2 | 96.5 | 95.9 | 95.5 | 94.7 | 94.2 | 94.2 | ||
# | 1861 | 1457 | 1132 | 861 | 648 | 460 | 240 | 59 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 2268 |
---|---|
Cumulative Months of Follow-Up | 93832 |
Number of Leads Active in Study | 519 |
Performance Notes For This Model

US Acute Lead Observations
Cardiac Perforation | 3 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 92 |
Failure to Capture | 60 |
Failure to Sense | 1 |
Impedance Out of Range | 9 |
Insulation Breach | 1 |
Lead Dislodgement | 205 |
Oversensing | 1 |
Unspecified Clinical Failure | 3 |

Product Characteristics
Lead Type | Left-Heart |
---|---|
Type and/or Fixation | Double Curve |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Left Ventricular Cardiac Vein |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | PVI |
---|---|
Lead Body Diameter | 4.0 FR |
Insulation | Polyurethane (outer); SI-polyimide (inner) |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of October 12, 2018