4196 Attain Ability
Left-Heart Pacing
4196 Attain Ability
ALL
Information

Device Survival Probability

4196_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr      
% Survival 98.0 97.3 96.6 95.9 95.7 95.0 94.7      
#   1843 1439 1104 829 617 400 152      

Product Surveillance Registry Results

Number of Leads Enrolled in Study 2250
Cumulative Months of Follow-Up 89803
Number of Leads Active in Study 569
Information 76

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 13
Failure to Capture 33
Impedance Out of Range 2
Insulation (not further defined) 1
Lead Dislodgement 21
Other 3
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 3
Conductor Fracture 2
Extra Cardiac Stimulation 89
Failure to Capture 58
Failure to Sense 1
Impedance Out of Range 9
Insulation Breach 1
Lead Dislodgement 205
Oversensing 1
Unspecified Clinical Failure 3
Information

Distribution Data

US Market Release2009-05-15
CE Approval Date2007-07-24
Registered USA Implants68025
Estimated Active USA Implants47379
US Market Release2009-05-15
CE Approval Date2007-07-24
Estimated WW Distribution187401
Information

USA Returned Product Analysis

Conductor Fracture 21
Insulation Breach 1
Crimp/Weld/Bond 0
Other 12
Information

WW Returned Product Analysis

Conductor Fracture 31
Insulation Breach 2
Crimp/Weld/Bond 0
Other 16
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Bipolar
More
Serial Number Prefix PVI
Lead Body Diameter 4.0 FR
Insulation Polyurethane (outer); SI-polyimide (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of January 31, 2018

www.medtronic.com/productperformance 601051