4195 Attain StarFix
Left-Heart Pacing
4195 Attain StarFix
ALL
Information

Device Survival Probability


4195_SURV
    1 yr 2 yr 3 yr 4 yr at 57 mo          
% Survival 99.2 98.6 98.3 98.3 98.3          
#   723 346 110 59 50          

System Longevity Study Results

Number of Leads Enrolled in Study 1231
Cumulative Months of Follow-Up 23520
Number of Leads Active in Study 963
Information 12

Qualifying Complications

Conductor Fracture 1
Extra Cardiac Stimulation 5
Failure to Capture 2
Insulation (not further defined) 1
Lead Dislodgement 3
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Extra Cardiac Stimulation 18
Failure to Capture 11
Impedance Out of Range 1
Lead Dislodgement 21
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release2008-08-15
CE Approval Date2005-05-13
Registered USA Implants14503
Estimated Active USA Implants11526
US Market Release2008-08-15
CE Approval Date2005-05-13
Estimated WW Distribution29888
Information

USA Returned Product Analysis

Conductor Fracture 1
Insulation Breach 1
Crimp/Weld/Bond 0
Other 4
Information

WW Returned Product Analysis

Conductor Fracture 2
Insulation Breach 1
Crimp/Weld/Bond 0
Other 14
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Deployable Lobes
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Cardiac vein
Pace/Sense Polarity Unipolar
More
Serial Number Prefix AAD
Lead Body Diameter 5.0 FR
Insulation Polyurethane
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050