4194 Attain OTW
Left-Heart Pacing
4194 Attain OTW
ALL
Information

Device Survival Probability


4194_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 75 mo      
% Survival 98.7 97.7 97.7 97.4 96.8 94.2 94.2      
#   934 611 372 222 141 68 52      

System Longevity Study Results

Number of Leads Enrolled in Study 1222
Cumulative Months of Follow-Up 36808
Number of Leads Active in Study 731
Information 28

Qualifying Complications

Extra Cardiac Stimulation 4
Failure to Capture 7
Insulation (ESC) 1
Insulation (not further defined) 2
Lead Dislodgement 14
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 2
Extra Cardiac Stimulation 28
Failure to Capture 34
Impedance Out of Range 6
Lead Dislodgement 114
Oversensing 2
Unspecified Clinical Failure 5
Information

Distribution Data

US Market Release2004-08-24
CE Approval Date2003-07-14
Registered USA Implants108173
Estimated Active USA Implants64822
US Market Release2004-08-24
CE Approval Date2003-07-14
Estimated WW Distribution187459
Information

USA Returned Product Analysis

Conductor Fracture 12
Insulation Breach 53
Crimp/Weld/Bond 0
Other 10
Information

WW Returned Product Analysis

Conductor Fracture 13
Insulation Breach 55
Crimp/Weld/Bond 0
Other 25
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Distal Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Cardiac vein
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LFG
Lead Body Diameter 6.0 FR
Insulation Polyurethane (55D) outer; silicone inner
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes