4194 Attain OTW
Left-Heart Pacing
4194 Attain OTW
ALL
Information

Device Survival Probability

4194_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr                    
% Survival 98.6 97.4 96.7 96.2 95.7 94.4 94.2 93.4 93.4 92.9 92.2                    
#   1370 1145 961 794 650 463 324 219 129 89 51                    

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1633
Cumulative Months of Follow-Up 84281
Number of Leads Active in Study 366
Information 64

Qualifying Complications

Conductor Fracture 2
Extra Cardiac Stimulation 11
Failure to Capture 19
Insulation (ESC) 1
Insulation (not further defined) 2
Lead Dislodgement 29
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 2
Extra Cardiac Stimulation 49
Failure to Capture 42
Impedance Out of Range 9
Lead Dislodgement 151
Oversensing 2
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release2004-08-24
CE Approval Date2003-07-14
Registered USA Implants114975
Estimated Active USA Implants52506
US Market Release2004-08-24
CE Approval Date2003-07-14
Estimated WW Distribution240393
Information

USA Returned Product Analysis

Conductor Fracture 35
Insulation Breach 134
Crimp/Weld/Bond 0
Other 2
Information

WW Returned Product Analysis

Conductor Fracture 41
Insulation Breach 139
Crimp/Weld/Bond 0
Other 3
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LFG
Lead Body Diameter 6.0 FR
Insulation Polyurethane (55D) outer; silicone inner
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of April 11, 2019