4193 Attain OTW
Left-Heart Pacing
4193 Attain OTW
ALL
Information

Device Survival Probability


4193_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 99 mo  
% Survival 95.9 95.0 94.3 94.3 93.8 93.3 93.3 92.3 92.3  
#   545 437 358 275 203 145 99 57 44  

System Longevity Study Results

Number of Leads Enrolled in Study 675
Cumulative Months of Follow-Up 30442
Number of Leads Active in Study 123
Information 38

Qualifying Complications

Conductor Fracture 1
Extra Cardiac Stimulation 8
Failure to Capture 12
Lead Dislodgement 14
Unspecified Clinical Failure 3
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Extra Cardiac Stimulation 16
Failure to Capture 11
Lead Dislodgement 44
Oversensing 1
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release2002-05-03
CE Approval Date2000-12-22
Registered USA Implants100672
Estimated Active USA Implants34954
US Market Release2002-05-03
CE Approval Date2000-12-22
Estimated WW Distribution174617
Information

USA Returned Product Analysis

Conductor Fracture 52
Insulation Breach 8
Crimp/Weld/Bond 0
Other 51
Information

WW Returned Product Analysis

Conductor Fracture 58
Insulation Breach 8
Crimp/Weld/Bond 0
Other 108
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Distal Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Cardiac vein
Pace/Sense Polarity Unipolar
More
Serial Number Prefix BAA
Lead Body Diameter 4.0 FR
Insulation Polyurethane (55D)
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050