4092 CapSure SP Novus
Pacing
4092 CapSure SP Novus
ALL
Information

Device Survival Probability


4092_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 117 mo
% Survival 98.9 98.8 98.7 98.4 98.1 97.7 97.7 97.7 97.7 97.7
#   929 831 734 637 553 449 320 206 117 47

System Longevity Study Results

Number of Leads Enrolled in Study 1147
Cumulative Months of Follow-Up 65701
Number of Leads Active in Study 293
Information 19

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 1
Failure to Capture 9
Impedance Out of Range 2
Lead Dislodgement 4
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Conductor Fracture 4
Failure to Capture 28
Impedance Out of Range 2
Insulation Breach 1
Lead Dislodgement 21
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release1998-09-17
CE Approval Date1998-04-15
Registered USA Implants176993
Estimated Active USA Implants80383
US Market Release1998-09-17
CE Approval Date1998-04-15
Estimated WW Distribution424896
Information

USA Returned Product Analysis

Conductor Fracture 9
Insulation Breach 48
Crimp/Weld/Bond 0
Other 3
Information

WW Returned Product Analysis

Conductor Fracture 9
Insulation Breach 52
Crimp/Weld/Bond 0
Other 10
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LEP
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050