4076
CapSureFix Novus
VENTRICULAR PLACEMENT
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | at 78 mo | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.8 | 99.8 | 99.7 | 99.7 | 99.0 | 99.0 | 99.0 | ||||
| # | 1087 | 829 | 534 | 353 | 200 | 76 | 44 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 1229 |
|---|---|
| Cumulative Months of Follow-Up | 46434 |
| Number of Leads Active in Study | 744 |
ATRIAL PLACEMENT
Device Survival Probability
4076_ATR_SURV
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | at 75 mo | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.8 | 99.6 | 99.5 | 99.5 | 99.5 | 98.9 | 98.9 | ||||
| # | 1471 | 1112 | 642 | 370 | 208 | 68 | 48 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 1662 |
|---|---|
| Cumulative Months of Follow-Up | 59076 |
| Number of Leads Active in Study | 1017 |
Performance Notes For This Model
US Acute Lead Observations
| Cardiac Perforation | 46 |
|---|---|
| Conductor Fracture | 4 |
| Extra Cardiac Stimulation | 8 |
| Failure to Capture | 63 |
| Failure to Sense | 14 |
| Impedance Out of Range | 11 |
| Insulation Breach | 1 |
| Lead Dislodgement | 116 |
| Oversensing | 6 |
| Unspecified Clinical Failure | 12 |
Product Characteristics
| Lead Type | Pacing |
|---|---|
| Type and/or Fixation | Active Screw In |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Atrium or Right Ventricle |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | BBL |
|---|---|
| Lead Body Diameter | 5.7 FR |
| Insulation | Polyurethane/Silicone |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Yes |
Data as of March 12, 2013