4076 CapSureFix Novus
Pacing
4076 CapSureFix Novus
VENTRICULAR PLACEMENT
Information

Device Survival Probability


4076_VEN_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 78 mo      
% Survival 99.8 99.8 99.7 99.7 99.0 99.0 99.0      
#   1087 829 534 353 200 76 44      

System Longevity Study Results

Number of Leads Enrolled in Study 1229
Cumulative Months of Follow-Up 46434
Number of Leads Active in Study 744
Information 5

Qualifying Complications

Extra Cardiac Stimulation 1
Failure to Capture 3
Impedance Out of Range 1
ATRIAL PLACEMENT
Information

Device Survival Probability


4076_ATR_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 75 mo      
% Survival 99.8 99.6 99.5 99.5 99.5 98.9 98.9      
#   1471 1112 642 370 208 68 48      

System Longevity Study Results

Number of Leads Enrolled in Study 1662
Cumulative Months of Follow-Up 59076
Number of Leads Active in Study 1017
Information 8

Qualifying Complications

Conductor Fracture 1
Failure to Capture 2
Failure to Sense 1
Insulation (not further defined) 1
Lead Dislodgement 3
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 46
Conductor Fracture 4
Extra Cardiac Stimulation 8
Failure to Capture 63
Failure to Sense 14
Impedance Out of Range 11
Insulation Breach 1
Lead Dislodgement 116
Oversensing 6
Unspecified Clinical Failure 12
Information

Distribution Data

US Market Release2004-02-25
CE Approval Date2004-06-14
Registered USA Implants432004
Estimated Active USA Implants332422
US Market Release2004-02-25
CE Approval Date2004-06-14
Estimated WW Distribution719858
Information

USA Returned Product Analysis

Conductor Fracture 33
Insulation Breach 33
Crimp/Weld/Bond 1
Other 24
Information

WW Returned Product Analysis

Conductor Fracture 39
Insulation Breach 36
Crimp/Weld/Bond 1
Other 37
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix BBL
Lead Body Diameter 5.7 FR
Insulation Polyurethane/Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013