4074
CapSure Sense
VENTRICULAR PLACEMENT
Device Survival Probability
4074_VEN_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | 16 yr | 17 yr | ||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.4 | 99.3 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 98.8 | 98.8 | 97.5 | 97.5 | 96.7 | 96.7 | 96.7 | 96.7 | |||||
# | 1029 | 878 | 737 | 624 | 481 | 393 | 337 | 294 | 252 | 203 | 171 | 151 | 121 | 117 | 111 | 84 | 52 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1193 |
---|---|
Cumulative Months of Follow-Up | 79817 |
Number of Leads Active in Study | 167 |
ATRIAL PLACEMENT
Device Survival Probability
4074_ATR_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | 16 yr | 17 yr | ||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | |||||
# | 214 | 205 | 198 | 183 | 167 | 158 | 148 | 136 | 126 | 117 | 110 | 107 | 98 | 94 | 88 | 73 | 51 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 227 |
---|---|
Cumulative Months of Follow-Up | 29196 |
Number of Leads Active in Study | 61 |
US Acute Lead Observations
Cardiac Perforation | 33 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 3 |
Failure to Capture | 183 |
Failure to Sense | 15 |
Impedance Out of Range | 6 |
Lead Dislodgement | 204 |
Oversensing | 8 |
Product Characteristics
Lead Type | Pacing |
---|---|
Type and/or Fixation | Tines |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | BBD |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) Silicone (inner) |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of December 1, 2023