4068
CapSureFix
VENTRICULAR PLACEMENT
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | at 147 mo | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.3 | 98.7 | 98.7 | 98.2 | 97.8 | 97.2 | 95.8 | 95.0 | 93.5 | 93.1 | 92.4 | 89.7 | 89.7 | |||||||||
| # | 1427 | 1219 | 1032 | 832 | 685 | 539 | 405 | 300 | 213 | 133 | 77 | 52 | 49 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 1799 |
|---|---|
| Cumulative Months of Follow-Up | 95888 |
| Number of Leads Active in Study | 162 |
ATRIAL PLACEMENT
Device Survival Probability
4068_ATR_SURV
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | at 165 mo | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 98.9 | 98.7 | 98.2 | 98.0 | 97.4 | 97.2 | 96.7 | 95.3 | 94.6 | 93.7 | 92.5 | 91.4 | 89.9 | 86.2 | ||||||||
| # | 1907 | 1637 | 1370 | 1116 | 897 | 737 | 577 | 472 | 358 | 256 | 189 | 135 | 84 | 43 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 2413 |
|---|---|
| Cumulative Months of Follow-Up | 133591 |
| Number of Leads Active in Study | 266 |
Performance Notes For This Model
US Acute Lead Observations
| Cardiac Perforation | 5 |
|---|---|
| Conductor Fracture | 3 |
| Extra Cardiac Stimulation | 1 |
| Failure to Capture | 23 |
| Failure to Sense | 5 |
| Impedance Out of Range | 2 |
| Insulation Breach | 1 |
| Lead Dislodgement | 31 |
| Unspecified Clinical Failure | 4 |
Product Characteristics
| Lead Type | Pacing |
|---|---|
| Type and/or Fixation | Active Screw In |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Atrium or Right Ventricle |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | LCE |
|---|---|
| Lead Body Diameter | 7.2 FR |
| Insulation | Polyurethane |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Yes |
Data as of March 12, 2013